| Designed to meet PQS E006/076 performance standards for cold?chain monitoring. Ideal for multi-stage distribution, bulk to last-mile monitoring. Displays real-time temperature readings and up to six alarms. Durable USB connector with gold-plated contacts for multiple insertions. Automatically generates PDF and Excel reports; no special software needed. Measurement range: -30°C to +60°C (-22°F to +140°F). Operating range: -30°C to +65°C (-22°F to +149°F). Storage range: 0°C to +40°C (32°F to +104°F). Accuracy: ±0.5°C (-20°C to +40°C), ±0.7°C outside this range. Resolution: 0.1°C/F, displayed to one decimal place. Sensor reaction time: <7 minutes (T90) in moving air. Recording capacity: 16,129 readings (56 days @ 5min, 168 days @ 15min). Sampling interval configurable from 30 sec to 18 hours. | Logging start: push-button or scheduled date/time. Recording indicated by “REC” symbol on LCD. Download full memory to PDF in <30 sec; LTD file in <10 sec via LogTag Analyzer. IP64 with USB cap for dust and splash protection. Power: UTRID-16R replaceable CR2032 or UTRID-16 fixed CR2450 battery. Battery life: 1–2 years depending on model and usage. Built-in real-time clock, ±25ppm accuracy at 25°C. USB 2.0 A-type plug for fast connection. Supports PDF Reader and LogTag Analyzer 2.8+. Multi-language support: English, Spanish, French, Russian, German. Dimensions: 93 x 54.5 x 8.6 mm; Weight: 39g. Polycarbonate case for durability. |